One-Step, High-Purity EV Isolation with Composite Columns

Writer: Upper Biotech
Upper Biotech
Nov 5
3 min read

Updated: 1 day ago

Introduction: the purity challenge in EV isolation


EVs carry proteins, nucleic acids and lipids that reflect their cell of origin, which makes them attractive as biomarker sources and therapeutic vehicles. Isolating them cleanly from complex biofluids is still a major challenge.


Conventional SEC preserves EV integrity and removes most soluble protein, but it cannot distinguish particles of similar size. In plasma and serum, lipoproteins (LDL, VLDL, chylomicrons) outnumber EVs by orders of magnitude and overlap them in size, so they co-elute.


For proteomics and LC-MS, these lipoprotein contaminants (ApoB, ApoA-I, ApoE and others) dominate the spectra and mask low-abundance EV proteins. Removing them usually means adding a second purification step, such as density gradient or affinity capture, which costs time and lowers recovery.


The Composite series is designed to solve this in one step.


Principle of the Composite columns


Composite columns integrate two separation mechanisms in one bed:

  1. Size exclusion. EVs are large and are excluded from the resin pores, eluting early in the void-volume fractions. Small soluble proteins such as albumin enter the pores and elute later.

  2. Ligand-activated core-shell capture. The resin core carries ligands that provide hydrophobic and ion-exchange interactions. Lipoproteins and other contaminants that would otherwise co-elute with EVs are bound and retained, while EVs pass through.


The result is an EV fraction that is depleted of both soluble proteins and lipoproteins, collected in the same simple gravity or automated workflow as standard SEC, with no extra steps.



Materials and methods


Columns and instrument

  • Composite series columns, Standard format or Maxi format

  • For comparison: Y series and P series SEC columns in the same formats


Samples

Human Plasma


Results

Lipoprotein and protein depletion across column series

Removal Rate (%)

Y series

P series

Composite series

Fraction

ApoA-I

ApoB

HSA

ApoA-I

ApoB

HSA

ApoA-I

ApoB

HSA

6

>99%

100.000%

99.999%

>99%

100.000%

99.998%

>99%

>99%

99.999%

7

>99%

96.923%

99.990%

>99%

93.846%

99.989%

>99%

>99%

99.996%

8

>99%

80.000%

99.967%

>99%

87.692%

99.982%

>99%

>99%

99.983%

9

>99%

80.000%

99.960%

>99%

87.692%

99.982%

>99%

>99%

99.935%

All three series remove essentially all albumin and HDL. The difference lies in LDL: because LDL overlaps EVs in size, pure SEC can only partly remove it. The Composite series adds ligand-based capture and depletes >99% of LDL (ApoB), the main contaminant that limits plasma and serum EV proteomics.



  1. Purity: Composite achieved the highest particle-to-protein ratio (721 ×10⁶ particles/µg), 3.0-fold higher than Y series and 2.4-fold higher than P series.

  2. Protein removal: Composite eluate contained only 0.143 mg/mL protein, 69% lower than Y series and 45% lower than P series.

  3. Particle recovery (NTA): Composite yield (103 ×10⁹/mL) was comparable to Y series (92%) and 32% higher than P series.


Overall, the Composite column improves purity mainly by removing more host-cell protein, without sacrificing particle recovery, outperforming both Y and P series controls.


Choosing the right column




Composite series

P series

Y series

Focus

Maximum purity

Balanced recovery and purity

Maximum recovery

Separation

SEC + hydrophobic + ion-exchange

SEC

SEC

Yield

10^9 - 10^11 particles/mL

10^9 - 10^11 particles/mL

10^10 - 10^12 particles/mL

Best for

Complex, protein-rich samples (plasma, serum)

Routine EV isolation

Low-EV or precious samples

Typical downstream use

Proteomics, LC-MS

Protein and RNA analysis

General EV characterisation

Use the Composite series when lipoprotein carry-over would compromise your results, particularly for plasma or serum proteomics and LC-MS-based biomarker discovery.


Conclusion


The Composite series brings two separation mechanisms into one column, a combination rarely found in commercial EV isolation products. It keeps the speed, gentleness and simplicity of SEC while adding the selectivity needed to remove lipoproteins. With >99% LDL (ApoB) depletion in a single step, it gives researchers working with plasma, serum and other complex samples a cleaner starting material for proteomics and LC-MS, without a second purification step.

 
 